NBRC RPFT - Registry Examination for Advanced Pulmonary Function Technologists Certification Examination Exam

Page: 1 / 1
Question #1 (Topic: Demo Questions)

When using a variable pressure plethysmograph, a leak in the door seal that develops subsequent to calibration will cause underestimation of

A.

R aw

B.

V TG

C.

Alveolar pressure


D.

vital capacity

Correct Answer: C
Explanation not available for this question.
Question #2 (Topic: Demo Questions)

A pulmonary function technologist is asked to select gas concentrations to simulate pediatric patients' exhaled concentrations during a D Lco simu - lation. Which of the following F E CO concentrations should the technologist select?

A.

0.200


B.

0.300


C.

0.400


D.

0.100

Correct Answer: B
Explanation not available for this question.
Question #3 (Topic: Demo Questions)

Results of two blood gas samples drawn from the same patient, 30 minutes apart, are shown below:

Which of the following is the most likely explanation for these differences?

A.

The patient's minute ventilation has increased.


B.

The patient has developed a pneumothorax.


C.

The 10:30 blood sample has been contaminated with air.


D.

The 10:30 blood sample contains an excess of heparin solution.

Correct Answer: A
Explanation not available for this question.
Question #4 (Topic: Demo Questions)

A pulmonary function technologist is performing quality control on a nebulizer used in the 5-breath dosimeter bronchial challenge. The target output of the device is 0.09 mL, plus or minus 10%. After 10 actuations, the nebulizer output was 75 µL with a 2.0 mL initial saline dose in the nebulizer. The technologist should

A.

Open the vent before starting the bronchial challenge.


B.

Add an exhalation filter and proceed with testing patients.


C.

Clean and reevaluate this nebulizer.


D.

Accept the results and begin using the device.

Correct Answer: D
Explanation not available for this question.
Question #5 (Topic: Demo Questions)

The following values are reported at maximum effort for a 50-year-old, 70-kg (154-lb) male with significant coronary artery disease during ergometer stress testing. Which of the following is most likely an error?

A.

workload 200 watts


B.

V E 65L/min


C.

HR145/min


D.

RER1.2

Correct Answer: A
Explanation not available for this question.
Download Exam
Page: 1 / 1
Next Page